Overview
SURETEMP PLUS 692 Thermometer
Advance your temperature routine with the SureTemp Plus 692 handheld thermometer featuring technology that sets the standard for speed and accuracy.
EASY
Easy-to-read backlit LCD display with user-selectable icon for identification of thermometer mode.
FAST AND ACCURATE
Approximately 4-6 second oral readings, 10-15 second for pediatric and adult axillary rectal readings.
ADVANCED TECHNOLOGY
Pulse timer button/monitor mode and recall button of last temperature taken.
SECURE
Protect against theft with location ID on the display.
SAFE
Interchangeable probe with push-button probe cover ejection and removable well for easy cleaning.
VERSATILE
Mode button for selecting oral, pediatric axillary or adult axillary.
Education & Documentation
Get in the know to get the most value out of your solution.
Product Documentation
-
Customer Letter
keyboard_arrow_down9600 Calibration Tester Obsolescence Letter, Customer Letter
-
Customer Service Bulletin
keyboard_arrow_downCleaning and Disinfection Recommendation, Customer Service Bulletin
-
Installation Guide
keyboard_arrow_downWelch Allyn SureTemp Plus Wall Holder, User Instructions
-
Quick Reference Guide
keyboard_arrow_downNormal Temperature Range, Quick Reference Guide
-
Service Manual
keyboard_arrow_downWelch Allyn SureTemp Plus 690 and 692 Series, Service Manual
-
User Manual
keyboard_arrow_downWelch Allyn SureTemp Plus Series 690/692, User Manual
-
User Manual (Previous Versions)
keyboard_arrow_downWelch Allyn SureTemp Plus Model 690 and 692 Thermometers, Service Manual
Frequently Asked Questions
Expand all-
What does the 5.0 error alert (e5.0) mean on my Welch Allyn SureTemp Plus 690/692 thermometer?
keyboard_arrow_downThis means the heater is not working properly. Try a new temperature probe; if the error clears, you may continue to use your unit. If there error does not clear, please contact Hillrom Customer Support for more assistance. -
My Welch Allyn SureTemp Plus 690/692 is stuck in axillary temperature mode. How do I change the default from axillary to oral temperature mode?
keyboard_arrow_downIn order to to do this, you must first be in Biotech Mode, then:
- Place the thermometer in the wall holder (Model 692 Only)
- Press and hold the Recall button, while removing the probe from the probe well. Once the probe is removed, release the Recall button.
- Press and release the Recall button until Default Mode is displayed in the upper left corner of screen
- To change between the Oral, Adult Axillary, and Pediatric Axillary modes, press and release the Mode Selection button. Observe that the display changes between ORAL, ADULT AXILLARY and PED AXILLARY when the Mode button is pressed.
- Replace the probe into the probe well to save default setting, then exit the Biotech mode. This will return the unit to normal operation.
-
My Welch Allyn SureTemp Plus 690 thermometer has an oral/axillary probe with part number 02692-000 or 02692-100 written on it, but my distributor says this is not a valid part number. What should I use to order a replacement probe?
keyboard_arrow_downWhile a Welch Allyn SureTemp Plus oral/axillary replacement probe alone is not available for purchase, you can get one as part of a kit that also includes a new probe well. Both parts should be replaced at the same time per device maintenance. The part number for this Welch Allyn SureTemp Plus Thermometer probe and well kit with a 4' coil cord is 02893-000; with a 9' coil cord, the part number is 02893-100. You can order this through your distributor.
-
My Welch Allyn SureTemp Plus 690/692 thermometer is reading in Celsius. How do I change the default setting to Fahrenheit?
keyboard_arrow_downIn order to to do this, you must first be in Biotech Mode, then:
- Place the thermometer in the wall holder (Model 692 Only).
- Press and hold the Recall button, while removing the probe from the probe well. Once the probe is removed, release the Recall button.
- Press and release the Recall button to move sequentially through the various program categories until you see “Units Default” in the upper left hand corner of the screen.
- Change default temperature mode to Celsius or Fahrenheit by pressing and releasing the C/F button.
- Replace the probe into the probe well and the device will go into Biotech mode and return to normal operation.
-
What is the difference between your Welch Allyn SureTemp Plus 690 and Welch Allyn SureTemp Plus 692 thermometers?
keyboard_arrow_downThe Welch Allyn SureTemp Plus 692 Thermometer contains the same features as the Welch Allyn SureTemp Plus 690 Thermometer, but also includes a wall mounting bracket, 60-second timer for counting pulses and respirations, backlit display, configurable on-screen labeling for instrument identification and electronic security options to help prevent theft. See our SureTemp Plus Brochure for more information
-
I'm getting an error code on my Welch Allyn SureTemp Plus 690 or 692 thermometer. What do I do resolve the error?
keyboard_arrow_downPlease refer to your Welch Allyn SureTemp Plus 690 or Welch Allyn SureTemp Plus 692 thermometer User Manual for an explanation of error codes.